kVp is the controlling factor for:
Improper tube, object and image receptor alignment with no c…
Improper tube, object and image receptor alignment with no change in SID will result in:
Pertaining to the diagram, each of the lines 1 – 4 represent…
Pertaining to the diagram, each of the lines 1 – 4 represents the intensity change due to filtration at different kVp levels.The greatest amount of attenuation is seen for valuse represented by:
If an AP thoracic spine is performed with a wedge compensato…
If an AP thoracic spine is performed with a wedge compensatory filter, the thick end of the filter should be placed over the:
The use of an x-ray tube with a small anode angle results in…
The use of an x-ray tube with a small anode angle results in: (Select all that apply)
Which generator produces the least amount of IR exposure per…
Which generator produces the least amount of IR exposure per mAs?
Which combination of variables will result in a radiograph w…
Which combination of variables will result in a radiograph with the lowest spatial resolution?
What is the main advantage of the use of a moving grid over…
What is the main advantage of the use of a moving grid over a stationary grid?
When the distance from the x-ray source decreases, the patie…
When the distance from the x-ray source decreases, the patient dose will:
Penumbra is ____ when the field size is ____ .
Penumbra is ____ when the field size is ____ .