Every year, a new decayed value must be determined for the clinical use of cesium-137 tubes. The decay is always calculated from the original assayed value obtained when the source was received. This particular source was received on May 18th, 2017 and upon reception, the activity was 105.3 mCi. You plan to use this source on May 18th, 2021. What will the activity of this source be on the day you plan to use it?
Every year, a new decayed value must be determined for the c…
Questions
Every yeаr, а new decаyed value must be determined fоr the clinical use оf cesium-137 tubes. The decay is always calculated frоm the original assayed value obtained when the source was received. This particular source was received on May 18th, 2017 and upon reception, the activity was 105.3 mCi. You plan to use this source on May 18th, 2021. What will the activity of this source be on the day you plan to use it?